Diagnostic accuracy of a home blood pressure monitor to detect atrial fibrillation

J Hum Hypertens. 2009 Oct;23(10):654-8. doi: 10.1038/jhh.2009.5. Epub 2009 Feb 12.

Abstract

Atrial fibrillation (AF) is the most common sustained arrhythmia and is associated with an increased long-term risk of stroke. A screening test for early diagnosis has the potential to prevent AF-related strokes. This study assessed the diagnostic accuracy of an automated device for self-home blood pressure (BP) monitoring, which implements an algorithm for AF detection. A modified, automated oscillometric device for self-home BP monitoring (Microlife BPA100 Plus, Microlife, Heerbrugg, Switzerland) with an AF detector was used to carry out triplicate BP measurements in subjects with sinus rhythm, AF and non-AF arrhythmias. During each BP measurement, the electrocardiogram (ECG) was recorded simultaneously. A total of 217 simultaneous BP measurements and ECG recordings were obtained from 73 subjects. Twenty-seven subjects (37%) had AF, 23 (31%) non-AF arrhythmias and 23 (31%) had sinus rhythm. A single measurement had 93% sensitivity and 89% specificity for detecting AF. For two measurements, in which one of them was required to detect AF, the sensitivity was 100% and specificity 76%, whereas for three measurements, in which two of them were required to detect AF, the sensitivity was 100% and specificity 89% (kappa=0.86 for an agreement with ECG). Using the latter approach, there were five false positive cases all having irregularities in approximately 50% of the heartbeats. In patients with tachyarrhythmia, the device underestimated heart rate. These data suggest that an electronic device for self-home BP monitoring, which implements an algorithm for AF diagnosis has an excellent diagnostic accuracy and might, therefore, be used as a reliable screening test for the early diagnosis.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Algorithms
  • Atrial Fibrillation / diagnosis*
  • Atrial Fibrillation / physiopathology
  • Blood Pressure Monitoring, Ambulatory / instrumentation*
  • Blood Pressure Monitors*
  • Blood Pressure*
  • Early Diagnosis
  • Electrocardiography
  • False Positive Reactions
  • Female
  • Heart Rate
  • Humans
  • Male
  • Middle Aged
  • Oscillometry
  • Predictive Value of Tests
  • Prospective Studies
  • Reproducibility of Results
  • Sensitivity and Specificity