Abstract
Commercial devices for tidal breathing measurements in newborns allow only short-term measurements, due to the high apparatus dead space of the face mask and pneumotachometer. The flow-through technique (FTT) minimizes the dead space by a background flow, thereby allowing long-term measurements. The aim of this study was to investigate the comparability of tidal breathing parameters using both techniques.
Paired measurements of tidal breathing were performed in 86 sleeping infants (median (range) body weight 2.8 kg (1.9–5.3 kg), age 65 days (3–150 days)), using the FTT and SensorMedics 2600 (SM 2600).
There was a significant bias (p<0.001) in all tidal breathing parameters. Compared with the FTT, increases (95% confidence interval (CI)) in tidal volume (VT), respiratory frequency (fR), and minute ventilation (V′E) were 0.74 (0.5–1.0) mL·kg−1, 9.0 (6.9–11.2)·min−1 and 92 (74–109) mL·min−1·kg−1 when measured with the SM 2600, representing average increases of 13, 17 and 30%, respectively, in response to the added dead space. By contrast, time to peak tidal expiratory flow as a proportion of expiratory time (tPTEF/tE) was changed by −0.09 (−0.11–−0.08). The mean (95% CI) change in tPTEF/tE of −54 (−62–−45)%, when measured in infants by the SM 2600, was remarkably similar to that observed during in vitro validation studies (−59 (−73–−44)%), suggesting that the discrepancies in timing parameters may be largely attributable to differences in signal processing.
In conclusion, differences in measurement technique and precision of the devices used can result in significant differences in tidal breathing parameters. This may impede the comparison of results within and between infants and the clinical interpretation of tidal breathing measurements in newborns.
This study was supported by the German Ministry for Research and Technology, project “Perinatal Lung” (grant 01 ZZ 9511) and Deutsche Forschungsgemeinschaft (Schm 1160/1‐2).
Most methods for respiratory function testing (RFT) in newborns currently require an accurate measurement of ventilation and correct recognition of the beginning of each inspiration and expiration. With the exception of forced expiratory manoeuvres, ventilatory measurements in infants are most commonly performed during tidal breathing 1. While both respiratory inductance 2 and “face out” body plethysmography 3 have been used to record tidal breathing without any facial attachments, thereby eliminating all external dead space, the former only provides quantitatively accurate measurements if measurement conditions remain very stable 4, while the latter is too expensive and cumbersome for routine bedside application 5. Consequently, direct measurement of airflow at the airway opening using a pneumotachometer (PNT) or similar flow meter connected to a face mask, has remained the most popular option for quantitative measurements.
Unfortunately, while being very simple, this measurement technique influences the very parameters it is designed to measure, respiration being stimulated by irritation of the nucleus trigeminus during facial attachment of the mask 6, 7, the additional resistance of the pneumotachometer (RPNT) and, most importantly, the apparatus dead space (VD,app) 8. The latter is of particular concern in newborns due to their low absolute tidal volumes (5–8 mL·kg−1 body weight) 9. While recent technological advances have facilitated the development of low dead space flowmeters, there are limitations to how much the effective dead space of the mask (which often represents the greatest component of VD,app during tidal breathing measurements) can be reduced 10. Since the water displacement of the smallest available face mask appears to be in the order of 10 mL, about half of which will represent effective dead space once in situ, this will present a considerable challenge to spontaneously breathing newborns, especially those <3 kg. Consequently, the duration of RFTs in newborns is limited if the conventional technique is applied. However, representative recordings of tidal breathing often require a relatively long acquisition period due to the high variability of sleep state and respiratory signals, especially in preterm newborns. Ruttimann et al. 11 described a method for overcoming this problem in ventilated neonates, by using differential pneumotachography based on flow measurements in the inspiratory and expiratory limb of the respirator circuit. This technique was subsequently used for ventilatory measurements in intubated, spontaneously breathing infants 12. An adaptation of this technique for tidal breathing measurements using a face mask is shown in figure 1⇓. This technique was called the flow-through technique (FTT) 13 because the face mask and PNT are continuously and thoroughly rinsed by a constant background flow which virtually eliminates VD,app. In contrast to the conventional technique, the functional reduction of VD,app by FTT allows long-term tidal breathing measurements.
The aim of this study was to investigate the comparability of the most commonly measured tidal breathing parameters, using both techniques, in newborns.
Materials and methods
Subjects
Tidal breathing measurements were performed in 86 newborns (table 1⇓) over a 10-month period (November 1996–August 1997), in the respiratory function laboratory of the Dept of Neonatology at Humboldt University (Charité, Berlin, Germany). The measurements were part of a prospective follow-up study of infants requiring mechanical ventilation during the neonatal period. In addition to other lung function tests such as the measurement of lung volume 14, paired measurements of tidal breathing parameters were performed using the conventional method and the flow-through technique. Admission criteria to this study were: spontaneous ventilation, quiet sleep according to Prechtl 15, and written parental consent. Infants with major malformations, upper airway obstruction, acute respiratory infection, congenital heart disease or central nervous malfunction were not eligible for recruitment. This study was approved by the medical faculty (Charité) of the Humboldt University (protocol 54/92). Parents were given a full explanation of the tests and equipment used before their consent was obtained.
Equipment for tidal breathing measurements
For the conventional tidal breathing measurements, the SensorMedics 2600 (SM 2600; SensorMedics Inc., Anaheim, CA, USA) was used, as previously described 16. Ventilatory measurements were performed using transparent face masks for premature neonates (size 1) or infants (size 2) (Vital Signs Inc., Totowa, NJ, USA), connected to a PNT (flow range: 0–10 L·min−1, RPNT=0.5 kPa·L−1·s total resistance of the PNT defined by the back pressure at 5 L·min−1, Hans Rudolph Inc., Wyandotte, MI, USA). The VD,app of the face mask and PNT was 9 mL for mask size 1 and 18 mL for mask size 2, determined by water displacement using a head model. Integration of airflow to derive volume was performed at 256 Hz without any filtering of the raw signals 16. The calibration of the PNT, including the linearization procedure, was performed prior to each measurement, according to the recommendations of the manufacturer. Custom made equipment was used for the tidal breathing measurements using the FTT (fig. 1⇑). The inflow and outflow of the face mask were measured by two screen PNTs (linear range 0–240 L·min−1, RPNT=0.2 kPa·L−1·s, total resistance, defined by the back pressure at 5 L·min−1, Baby PNT Jaeger, Würzburg, Germany) and the infant's airflow was given by the difference between the two flow signals. The continuous background flow, which was generated by an air mixer and flow regulator, was fed into the inspiratory limb and the same face mask as used with the SensorMedics 2600 (SM 2600) via a modified y-piece. Flow was recorded at 200 Hz and filtered by an analogue Bessel filter of fourth order with a cut-off frequency of 48' Hz to avoid aliasing 17. The software Labview (Version 4.0.1., National Instruments, Austin, Texas, USA) was used for data acquisition and further signal processing. The in vitro measurement accuracy of the two devices was assessed as described previously, using a mechanical model pump to generate flow profiles that simulated those observed in infants 18.
Measurements in infants
Normally, a clean, dry newborn falls asleep in comfortable environmental conditions within about half an hour of feeding. This biological reaction and the infant's sleep-waking rhythm were used to determine timing of respiratory function tests. Most infants were studied during natural, quiet sleep, assessed by behavioural criteria 15, but 27 children (31%) were sedated with chloral hydrate (50 mg·kg−1) given orally 15–30 min before testing. Sedation was primarily necessary in older infants and for the additional lung function tests, such as measurement of lung volume. Parents were usually present during the RFT. For the measurements, sleeping infants were placed in a supine position with the neck in a neutral position, supported by a head seal. After the infant was asleep, tidal breathing measurements were performed, firstly with the FTT, followed immediately afterwards with the SM 2600. The duration of both measurements was normally 20–30 min, depending on the period of acclimatization to the face mask. Infants were continuously monitored throughout the whole procedure by pulse oximetry.
During the FTT, the constant background flow (V′const) was adjusted in such a way that it was always higher than the infant's peak tidal inspiratory flow 13. After a sufficient period of adaptation (5–20 min), tidal breathing was measured and the airtight seal of the mask on the infant's face was checked by continuous leak monitoring 19. The end of the adaptation period is commonly characterized by a more regular breathing pattern without any visible drift in breathing parameters. When using the FTT, on-line display of time based signals and flow/volume loops facilitate both recognition of the end of the period of adaptation and selection of suitable epochs, with low breath-to-breath variability and minimal artefacts for subsequent evaluation. The graphical display of instantaneous respiratory frequency over the last 60 breathing cycles also assists assessment of the stability of the tidal breathing parameters.
Depending on the variability of the breathing pattern, at least 10, but not more than 30 consecutive breaths were evaluated. Immediately afterwards, tidal breathing measurements were carried out with the SM 2600. An identically sized face mask connected to the Rudolph PNT was placed on the face. When using the SM 2600, single breathing cycles of the stored signals have to be selected by the investigator. For the current study, tidal breathing parameters were determined from 4–12 selected breathing cycles when the infant was breathing regularly.
Statistical methods
Mean±sd was calculated for all ventilatory parameters, together with the mean (95% confidence interval (CI)) of the within-subject differences in tidal breathing parameters between the two methods. The comparability of the two techniques was assessed using paired t‐tests and the statistical approach described by Bland and Altman 20. The software STATGRAPHICS (Version 4.0, Manugistics Inc., Rockville, Maryland, USA) was used for the statistical evaluations. A level of statistical significance of p<0.05 was accepted.
Results
The comparison of the tidal breathing parameters recorded with both methods are shown in table 2⇓, where statistically significant differences between techniques for all parameters investigated (p<0.001) are displayed. Thus, when compared with the FTT, tidal volume (VT), respiratory frequency (fR) and minute ventilation (V′E) were on average 13, 17 and 30% higher, respectively, when measured by the SM 2600. Even greater differences were found for parameters dependent on the shape of the flow signal or the tidal breathing flow/volume loop (TBFVL), with mean±sd time to peak tidal expiratory flow as a portion of expiratory time (tPTEF/tE) decreasing from 0.24±0.08 (FTT) to 0.15±0.08 (SM 2600), representing a mean (95% CI) change in tPTEF/tE of −54 (−62– −45)%. The parameter volume to peak tidal expiratory flow as a portion of tidal volume (VPTEF/VT), which describes the site of maximum expiratory flow on the TBFVL, and which is strongly correlated with tPTEF/tE (r=0.959 for SM 2600 and r=0.834 for FTT), was also found to have decreased (by about 39%) during measurements with the SM 2600. By contrast, the parameter expiratory flow when 25% of tidal volume remains in the lung (TEF25), which reflects the curvature of the end-expiratory limb of the TBFVL, was significantly higher (p<0.001) when measured with the SM 2600. Individual within-subject differences for VT, fR and tPTEF/tE are shown in figure 2⇓, plotted according to the Bland and Altman method 20. It can be seen that while there was considerable individual variability to the change in measurement technique, with wide limits of agreement between the methods, a significant bias was evident in all the measured parameters (table 2⇓).
Discussion
Relative advantages and limitations of the two methods
The conventional technique, which simply requires a face mask and low-dead space PNT, is widely used for RFT in newborns. The physical properties of the gas flow through the PNT (room air during inspiration, alveolar gas with body temperature and pressure saturated (BTPS) conditions during expiration) can be corrected 21. In contrast to the FTT, it can be used in conjunction with techniques requiring airway occlusions (e.g. plethysmography and passive respiratory mechanics 1). However, a crucial disadvantage of this method is the high apparatus dead space, which in newborns often exceeds the physiological dead space. The duration of measurements, therefore, needs to be as short as possible in such infants to avoid compromising alveolar gas exchange. This disadvantage can be overcome by using the FTT. The virtual elimination of VD,app by the background flow permits long−term measurements, the duration of which can readily be adapted to the prevailing measuring conditions (e.g. time required to reach a steady state) as well as to check the variability of the respiratory signals. In addition, the FTT has the advantage that the inspired oxygen fraction can be adjusted, tracer gases for other RFTs can be added 13, and it can easily be used in infants requiring continuous positive airway pressure 22. Furthermore, the FTT can be combined with oesophageal manometry 23 or forced oscillations to investigate lung mechanics during tidal breathing 24. The FTT also allows continuous assessment of air leaks between the face and the mask which is helpful for maintaining airtight placement of the face mask 19.
Potential disadvantages of the FTT include the higher technical expenditure and demands on both PNTs (e.g. linear characteristics required over a wider flow range to avoid any influence of the background flow on the measurement). In addition, as mentioned previously, the FTT cannot be used for all respiratory function tests, particularly those requiring airway occlusions. Nevertheless, it would appear to be the method of choice for tidal breathing measurements in newborns, especially those delivered prematurely.
Factors influencing tidal breathing measurements
The biggest difference between the techniques used in this study was with respect to the apparatus dead space. In contrast to the FTT, the VD,app of the SM 2600 (9 or 18 mL, depending on mask size selected) could have a significant influence on alveolar gas exchange and blood gases. Although the apparatus resistance of the SM 2600 was also higher than that of the FTT (0.5 versus 0.2 kPa·L−1·s respectively), both are considerably smaller than the infant's own resistance and it is, therefore, unlikely that this difference in resistive load between methods made any major contribution to the observed differences in the tidal breathing parameters. As shown in table 1⇑, body weight was less than 2,800 g in half of the infants studied, and in such infants the intrinsic resistance is likely to be nearly ten times higher that that of the apparatus, even when using the SM 2600 25. While the air stream from the background flow on the infant's face could potentially influence measured tidal breathing parameters when using the FTT, it has been previously shown that this flow has no significant influence on breathing pattern, provided infants are allowed to adapt for a sufficient period of time 26. All other factors that could potentially influence tidal breathing were the same for both techniques.
Ideally, order of measurements would have been randomised, whereas in this study the FTT was always applied first. This was to ensure that the two sets of measurements could be obtained as closely in time as possible without any interruption of infant's sleep. Had the SM 2600 been applied first in some infants, a much longer interval would have been required for “wash out” because, in contrast to FTT, measurements with the SM 2600 have a significant influence on both tidal breathing parameters 27 and blood gases. Such a prolonged waiting time (>30 min) would have hampered the comparability of the measured parameters and decreased the feasibility of completing the protocol in many of the infants. It has previously been shown that actual duration of measurements may also influence tidal breathing parameters 27. The FTT allows the investigator to wait until a steady state in the parameters has occurred, whereas during measurements with the SM 2600, the first regular breaths within the first 1–2 min after placement of the face mask have to be used due to the dead space ventilation.
Important differences also exist between the two devices in the estimation of breath-to-breath variability which precludes direct comparisons. For the SM 2600, this is based on user selection of breathing cycles with large amplitudes and minimal disturbances and deformations 28, which has a high risk of subjective bias 5. In contrast, consecutive evaluation of a sufficient number of breathing cycles was used in the FTT to ensure a more representative evaluation, as described previously 29. However, since RFTs were only performed in sleeping infants with a regular breathing pattern, this should not have been a source of discrepancy in this particular study.
Differences in the measured tidal breathing parameters
The increase in VT and fR seen during measurements in infants with the SM 2600 (table 2⇑), are in agreement with a previous study which found increases in the variables of about 25 and 20% respectively, within 2–3 min of mask placement when using the same equipment 27. This was primarily attributed to carbon dioxide (CO2) rebreathing due to the high apparatus dead space. The fact that tidal breathing parameters measured with the conventional technique are so dependent on the time of measurement after placement of the face mask, increases the difficulties in attempting to standardize such measurements.
The parameter TEF25 described the end-expiratory course of the TBFVL and was significantly higher when measured with the SM 2600 (table 2⇑). This may reflect the fact that such flows were measured at a higher lung volume due to the increased VT per se and/or dynamic elevation of the end expiratory level, secondary to the increased fR and decreased tE with this device 30, 31. In this study, the greatest differences between measurement techniques were with respect to tPTEF/tE and VPTEF/VT, with values being reduced by up to 50% when compared with the FTT. It is unlikely that such differences can simply be attributed to the physiological changes in VT and fR induced by the added dead space, particularly since similar errors can be identified during in vitro simulations 18. During in vitro validation studies, no differences were observed in the raw flow signal obtained from the two devices, nor between values calculated automatically or manually from the FTT 18. Similarly, the deviations between the FTT and SM 2600 were <3% for VT, fR and V′E. However, when compared either with manually calculated values or those derived automatically from the FFT, there was a significant underestimation of shape-dependent parameters such tPTEF/tE derived by the SM 2600, with errors being up to 60% when flows simulating those observed in preterm neonates were applied. These errors appear to result from differences in signal processing and can only be detected if appropriate nonsinusoidal flow profiles, representing those seen in infants, are used to evaluate equipment 18.
Both tPTEF/tE and VPTEF/VT are not only critically dependent on the accurate determination of PTEF, but on the accurate determination of zero flow and hence, start of expiration. In the FTT, which was developed recently, considerable efforts were made to ensure correct breath detection even in the smallest infants, by using a flow-threshold algorithm with an additional plausibility check for the zero crossings that were detected 32. Further validation was performed by comparing results that were derived automatically with manual calculations of the time-based trace and TBFVL. Unfortunately, such an approach was not feasible with the SM 2600, since this was developed over 15 yrs ago, before recent technological advances in equipment and software. Since the algorithms used by the SM 2600 for signal evaluation were not available, and manual evaluation of the graphical presentation was not feasible, detailed analysis of the source of the discrepancies was not possible. The results from this study suggest that equipment for infant RFTs needs to be carefully checked using a suitable in vitro model before clinical application 18, 33, 34. Sinusoidal model pumps, which are commonly used to investigate the accuracy of devices for infant RFT, are not necessarily appropriate. Instead, model pumps with adjustable flow profiles covering the relevant measuring range, should be used in such assessments 18.
Clinical implications
It is a fundamental principle of all physiological measurements that measuring procedures should have no or minimal influence on the parameters being measured, and that different techniques and equipment should lead to the same results. However, this demand is extremely difficult to achieve when measuring tidal breathing in newborns. The results from this study suggest that tidal breathing parameters measured with the FTT differ significantly from those collected with conventional systems that have an appreciable dead space. Hopefully, in the near future, precise flowmeters with negligible flow resistance (e.g. anemometers 35 or ultrasonic sensors 36) will become routinely available, so that, in addition to the elimination of apparatus dead space by the FTT, the resistive load can also minimized.
One important message from this study is that some differences (especially in shape-dependent parameters like time to peak tidal expiratory flow as a proportion of tidal expiratory time) are so large that there is currently no comparability in the measured tidal breathing parameters collected using different devices, and diagnostic misinterpretations are, therefore, likely. The fact that tidal breathing parameters in newborns are currently so method-dependent makes it difficult to compare data between different laboratories and impossible to pool data for a meta-analysis (e.g. for establishment of reference ranges). Furthermore, reference values which have been determined with any one particular system 37 are unlikely to be of relevance when using other measurement techniques or equipment. Thus, there is an urgent need to standardize the technical requirements for respiratory function testing in newborns, with respect to both equipment and software.
Acknowledgments
The authors thank M. Schmidt for support in the development of the software used and D. Windstetter and S. Schmidt for their assistance in respiratory function testing.