Abstract
Multiple breath nitrogen washout (MBNW) quantifies ventilation heterogeneity. Two distinct protocols are currently used for MBNW testing: “controlled breathing”, with targeted tidal volume (VT) and respiratory rate (RR); and “free breathing”, with no constraints on breathing pattern. Indices derived from the two protocols (functional residual capacity (FRC), lung clearance index (LCI), Scond, Sacin) have not been directly compared in adults. We aimed to determine whether MBNW indices are comparable between protocols, to identify factors underlying any between-protocol differences and to determine the between-session variabilities of each protocol.
We performed MBNW testing by both protocols in 27 healthy adult volunteers, applying the currently proposed correction for VT to Scond and Sacin derived from free breathing. To establish between-session variability, we repeated testing in 15 volunteers within 3 months.
While FRC was comparable between controlled versus free breathing 2.92 (0.86) versus 2.94 (0.89) L, p=0.75), indices of ventilation heterogeneity derived from the two protocols were not, with poor correlation for Scond (r=0.18, p=0.36) and significant bias for Sacin (0.057 (0.021) L−1 versus 0.085 (0.038) L−1, p=0.0004). Between-protocol differences in Sacin were related to differences in the breathing pattern, i.e. VT (p=0.004) and RR (p=0.01), rather than FRC. FRC and LCI showed good between-session repeatability, but Scond and Sacin from free breathing showed poor repeatability with wide limits of agreement.
These findings have implications for the ongoing clinical implementation of MBNW, as they demonstrate that Scond and Sacin from free breathing, despite VT correction, are not equivalent to the controlled breathing protocol. The poor between-session repeatability of Scond during free breathing may limit its clinical utility.
Abstract
Phase 3 slopes indices derived from “free breathing” and “controlled breathing” MBNW protocols are not comparable, and differences are related to breathing patterns. These findings have implications for the ongoing clinical implementation of MBNW. https://bit.ly/35oQYnW
Introduction
Increased ventilation heterogeneity is a characteristic physiological abnormality in respiratory diseases such as asthma and COPD [1, 2]. The multiple breath nitrogen washout (MBNW) test is an increasingly available method of quantifying ventilation heterogeneity. MBNW is conducted by breathing 100% oxygen (O2) which then “washes out” the resident nitrogen (N2) in the lung. Analysis of the exhaled N2 concentration versus exhaled volume of each breath then allows calculation of a number of parameters: a global measure of heterogeneity known as the lung clearance index (LCI); heterogeneity arising predominantly within the conducting airways (Scond); heterogeneity arising predominantly in the more distal/intra-acinar airways (Sacin); and functional residual capacity (FRC).
Two distinct breathing protocols are currently used for MBNW testing. One is a “controlled breathing” protocol whereby the patient maintains a consistent tidal volume (VT) (∼1 L) and respiratory rate (RR) (8–12 breaths·min−1) [3] through the use of a visual incentive screen and real-time feedback from the test operator. These breathing constraints are applied to all individuals equally, regardless of lung size; consequently, a given patient's breathing pattern during the test may deviate significantly from their usual resting breathing pattern. The alternative is a “free breathing” protocol, with no constraints on VT or RR, i.e. the individual determines their own breathing depth and frequency. While this makes the method more suited to paediatric testing [3], variability in breathing patterns within or between individuals may have an impact on indices of ventilation heterogeneity [4–6] and their repeatability. A correction for lung size and breath-to-breath variation in VT has been proposed for this method [7], but this has not been formally validated against a controlled breathing approach. Standard reference equations are available for both methods to guide interpretation of MBNW indices, which account for factors such as age, height and sex [8–11], but little is known about their comparability.
Differences in how controlled and free breathing MBNW protocols are conducted and analysed may therefore limit the comparability of indices derived from them. A direct comparison of the protocols in children [12] showed significant differences in LCI and Scond; the authors concluded that this was due to a reduction in end-expiratory lung volume (EELV) induced by controlled breathing. However, the results suggested that other patient-related factors (e.g. lung size, body weight) may contribute to the discrepancies. Indices derived from the two protocols have not been directly compared in adults, nor have their respective between-session variabilities.
Therefore, in a sample of healthy adults, we aimed to determine: 1) whether controlled and free breathing MBNW protocols provide equivalent FRC and indices of ventilation heterogeneity; 2) the influence of patient-related factors (anthropometrics and/or breathing pattern) on any observable differences; and 3) if the choice of protocol influences between-session variability, which would allow us to interpret any between-protocol differences. Our overall hypothesis was that controlled and free breathing MBNW would produce comparable measurements of ventilation heterogeneity, with the secondary hypothesis that any differences would be, in part, due to patient-related factors.
Methods
Study overview
We recruited and studied volunteers aged ≥18 years in the respiratory function laboratories at two sites (Royal North Shore Hospital (RNSH) and Woolcock Institute of Medical Research, Sydney, Australia; see supplementary material for inclusion/exclusion criteria) from a convenience sample of volunteers (predominantly hospital staff) between April 2018 and November 2019. Our target sample size was n=25, comparable to the control group of a previous study [12]. The RNSH Human Research Ethics Committee approved the study (LNR/16/HAWKE/11). After obtaining written informed consent to be tested and for their data to be used for research purposes, participants first underwent standard (pre-bronchodilator) lung function testing to confirm they had normal lung function. After a period of rest of at least 10 min, participants underwent MBNW testing by both breathing protocols conducted in a randomised order. Each participant was invited to return for repeat testing, at their convenience, within the next 3 months. This follow-up testing was performed in an identical manner, including the order of MBNW protocol testing, for consistency.
Standard lung function testing
We measured spirometry and lung volumes according to American Thoracic Society/European Respiratory Society (ATS/ERS) quality criteria using a Jaeger MasterScreen PFT (Vyaire Medical GmbH, Hoechberg, Germany) device, with comparison to reference values [13, 14] for plethysmography.
MBNW testing
Detailed descriptions of the device and testing procedure are found in the supplementary material. After at least 10 min of rest for the participant, we conducted MBNW tests using the Exhalyzer D with Spiroware v3.1.6 (Eco Medics AG, Duernten, Switzerland).
Participants performed both controlled and free breathing protocols in succession, in a random order (as determined by a computer-based random number generator), according to current international consensus recommendations [3]. All trials started with a period of normal relaxed breathing on room air in order to establish a stable EELV. For the controlled breathing protocol, the operator then instructed the participant to breathe at a VT of 0.95–1.3 L and at RR 8–12/min with the use of the visual incentive screen within Spiroware (described by Verbanck et al. [15]); once the operator was satisfied that a stable breathing pattern and EELV had been achieved, they commenced the washout by switching the circuit to 100% O2, and the participant maintained the same breathing pattern for the duration of the washout phase. For the free breathing protocol, the operator instructed the participant to “continue to breathe in a normal relaxed manner” through both the pre-phase and washout phase, with the visual incentive screen switched off and without any additional coaching during the trial. Participants repeated the test until at least three technically acceptable trials with FRC values within ±10% of the mean were achieved for each protocol.
MBNW analysis
Full details are provided in the supplementary material. We analysed MBNW data using Spiroware software (v.3.1.6). While the test operator performed preliminary analysis on individual trials during the testing session, a single investigator re-analysed the data post hoc for all participants as a batch in order to ensure a consistent approach to analysis.
To differentiate between measurements made by the different protocols, we have subscripted all indices with CB or FB to indicate controlled and free breathing, respectively. Under the free breathing protocol, Scond and Sacin are adjusted for VT and are thus denoted in the literature and Spiroware software as Scond*VT and Sacin*VT; however, we refer to them here simply as ScondFB and SacinFB, respectively.
Statistical analyses
We compared FRC, LCI, Scond and Sacin measured by the two breathing protocols using paired t-tests and Pearson correlation. Additionally, we compared FRC from either protocol against the gold-standard FRCpleth. To investigate for bias, we generated Bland–Altman plots as the between-protocol difference (free breathing minus controlled breathing) versus the average, plotting the mean difference and 95% limits of agreement (95% limits of agreement (LOA)). We then performed linear regression of the difference versus average to determine any proportional bias. To examine the effects of various predictors (age, sex, height, BMI, mean RR from free breathing, mean VT from free breathing and FRCpleth) on between-protocol difference, we performed linear regression of the difference versus each predictor. To determine within-session variability, we calculated the coefficient of variation (CoV) from the three washout trials. To determine between-session variability, we calculated the difference (Visit 2 minus Visit 1) and 95% LOA separately for each protocol. We also report the between-session intra-class correlation coefficients (ICC), calculated using a two-way mixed effects ANOVA model based on absolute agreement, multiple measurements (k=3) [16]. We set statistical significance at p<0.05. Results are presented as mean (standard deviation (sd)) unless otherwise stated.
Results
Participant characteristics and breathing patterns
We studied 27 non-smoking, healthy volunteers (22 at the RNSH site, 5 at the WIMR site) with a median age of 34 (range 19–65) years and spirometry/lung volumes within the limits of normal (table 1). Since MBNW parameters are likely to be device- and protocol-specific [17, 18], we were unable to compare all measurements to a single set of reference equations. However, MBNW indices from the free breathing protocol were within the ranges of normal derived from the same testing device (supplementary table S1) [9].
Participant characteristics, lung function and multiple-breath nitrogen washout (MBNW) parameters
Compared to controlled breathing, free breathing produced smaller mean VT (mean (sd) difference −0.24 (0.33) L, p=0.0006), faster mean RR (2.14 (3.45) breaths·min−1, p=0.003), smaller minute ventilation (−2.4 (3.1) L·min−1, p<0.0001) and greater cumulative expired volume (CEV) (1.84 L, p=0.006), but similar washout times (p=0.17). Four out of the 27 participants required coaching prior to commencing the free breathing protocol due to inadequate phase III. The average proportions of washout breaths excluded from analysis due to inadequate phase III were 2.7% for controlled breathing and 3.9% for free breathing.
FRC was comparable between protocols and with plethysmography
There was no significant difference in mean FRC measured by either MBNW protocol (FRCCB 3.17 (0.98) versus FRCFB 3.18 (0.94), p=0.88). FRCCB and FRCFB were strongly correlated (r=0.94, p<0.0001; figure 1a). There was no evidence of bias in the absolute (figure 1b) or percentage difference (supplementary figure S1).
Functional residual capacity measured by controlled breathing (FRCCB) and free breathing (FRCFB) protocols. a) There was strong correlation between the protocols (r=0.96, p<0.0001). b) Bland–Altman plot showing good agreement between the protocols (mean difference (95% limits of agreement) −0.009 (−0.592, 0.555) L, p=0.75).
There were significant differences between mean FRCCB or FRCFB and mean FRCpleth (p=0.83 and p=0.86, respectively). Both FRCCB and FRCFB were strongly correlated with FRCpleth (r=0.84 and r=0.92, respectively, p<0.0001 for both; figure 2). Within-subject differences in FRC between the two MBNW protocols were not related to the individual's FRCpleth (p=0.55, supplementary figure S2-A). Similarly, within-subject differences in FRC between the protocols were not related to age, sex, height, mean VT or mean RR from free breathing, but there was a weak effect of BMI (p=0.04, supplementary figure S2-B).
Functional residual capacity measured by controlled breathing (FRCCB) and free breathing (FRCFB) protocols versus the gold-standard body plethysmography (FRCpleth). There was good correlation between FRC measured by both protocols and FRCpleth (r=0.86 and r=0.92, respectively, p<0.0001 for both). MBNW: multiple breath nitrogen washout.
LCI was comparable between protocols
Mean LCI was significantly lower with controlled breathing (LCICB 7.2 (0.58) versus LCIFB 7.55 (0.81), p=0.0004); however, the mean difference was small. There was a strong correlation between the protocols (r=0.84, p<0.0001; figure 3a). There was evidence of proportional bias (between-protocol difference increased with LCI, p=0.004; figure 3b).
Lung clearance index measured by controlled breathing (LCICB) and free breathing (LCIFB) protocols. a) There was strong correlation between the protocols (r=0.82, p<0.0001). b) Bland–Altman plot showing that free breathing produced a higher LCI compared to controlled breathing (mean difference (95% limits of agreement) 0.21 (−0.44, 0.87), p=0.003)). There was a trend towards proportional bias confirmed by linear regression (p=0.07).
Scond was poorly correlated between protocols
Overall, mean Scond measured by both protocols was not significantly different (ScondCB 0.017 (0.009) versus ScondFB 0.018 (0.01) L−1, p=0.74). However, there was no significant correlation between the protocols (r=0.18, p=0.36, figure 4a). The Bland–Altman plot revealed large variance in between-protocol differences and possible (but non-significant) proportional bias (p=0.45, figure 4b).
Ventilation heterogeneity in conducting airways measured by controlled breathing (ScondCB) and free breathing (ScondFB) protocols. a) There was no significant correlation between the protocols (r=0.16, p=0.44). b) Bland–Altman plot showing high between-protocol variability (mean difference (95% limits of agreement) 0.0002 (−0.03, 0.03 L−1, p=0.94)). There was no evidence of proportional bias, regardless of outliers (p=0.83).
Sacin showed significant between-protocol differences
Mean Sacin was significantly lower with controlled breathing (SacinCB 0.057 (0.021) L−1 versus SacinFB 0.085 (0.038) L−1, p=0.0004). The correlation between the protocols was borderline significant (r=0.37, p=0.06; figure 5a). The Bland–Altman plot revealed significant proportional bias (between-protocol difference increased with Sacin, p=0.002; figure 5b). Linear regression showed that within-subject differences in Sacin between the two protocols were related to the breathing pattern. Specifically, the greater the deviation in mean VT or mean RR between the protocols, the larger the discrepancy in Sacin (p=0.004 and p=0.01, respectively; figure 5c and d), such that participants who breathed shallower or faster during the free breathing protocol had greater apparent SacinFB.
Ventilation heterogeneity in distal/intra-acinar airways measured by controlled breathing (SacinCB) and free breathing (SacinFB) protocols. a) There was relatively poor correlation between the protocols (r=0.33, p=0.11). b) Bland–Altman plot showing that free breathing produced higher Sacin compared to controlled breathing (mean difference (95% limits of agreement) 0.03 (−0.045, 0.103) L−1, p<0.0005)). There was also significant proportional bias confirmed by linear regression (p=0.002). The between-protocol difference in Sacin (SacinFB–SacinCB) was predicted by the between-protocol differences in c) tidal volume (VTFB−VTCB, regression p=0.004) and d) respiratory rate (RRFB−RRCB, regression p=0.009). One participant was excluded from Sacin analyses due to negative value in one trial.
Breathing protocol influenced between-session variability of MBNW indices
Fifteen volunteers underwent repeat testing; the baseline (Visit 1) characteristics of this subgroup were similar to those of the group as a whole, except for a narrower age range (table 1). The mean (sd) time between sessions was 5.9 (3.3) weeks. Within-session and between-session variability measures for both protocols are presented in table 2. There were no differences seen in within-session CoV between protocols in FRC (p=0.193) or LCI (p=0.240). In terms of between-session variability, the free breathing protocol showed relatively greater variability in LCI, Scond and Sacin, as indicated by numerically greater mean differences and wider 95% LOAs, with very poor between-session ICC seen in Scond and Sacin. The controlled breathing protocol showed wider 95% LOA for FRC. Bland–Altman plots did not suggest any proportional bias for either protocol (supplementary figures S3 and S4).
Within- and between-session variability for the controlled and free breathing protocols
Discussion
Summary of results
In this study in healthy adults comparing two commonly used MBNW breathing protocols, we found that: 1) FRC was comparable between the two protocols, however indices of ventilation heterogeneity based on phase III slopes (SIII) were not – Scond was poorly correlated between the two protocols and Sacin was systematically higher under the free breathing protocol, whereas LCI was only marginally higher; 2) there was greater between-protocol discrepancy in Sacin in subjects whose free-breathing pattern deviated from that of the volume- and frequency-controlled protocol; and 3) Scond and Sacin exhibited poorer between-session repeatability under the free breathing protocol.
Lack of discrepancy in FRC
Yammine et al. [12] demonstrated in children that Scond and LCI obtained from a controlled breathing protocol were higher than when obtained by free breathing and that the discrepancy was related to a decrease in FRC during controlled breathing. Thus, the authors argued that a controlled breathing protocol may overestimate heterogeneity by inadvertently causing individuals to change the EELV at which they would otherwise normally breathe. In children, a 1 L VT represents a significant proportion of total lung capacity; the high VT may increase ventilation heterogeneity due to the recruitment of normally non-ventilated lung units [19], or due to closure of lung units in the dependent regions if the child expires to below their normal EELV [20], or some combination of both.
Our current findings in healthy adults do not support the argument that FRC is altered with controlled breathing, as we showed that FRC estimated by the two protocols was in fact comparable, with no systematic bias. Furthermore, both were comparable to the “gold standard”, i.e. FRCpleth. A number of factors may explain the differences between our findings and those of Yammine et al. For example, the effect of variation in FRC may be relatively mild in the adults in our study compared to that in children, and thus insufficient to cause significant differences in heterogeneity. We observed that the variability in between-protocol differences (assessed by the LOA) in FRC were comparable to the between-session variabilities of either protocol, though interestingly between-session LOA was numerically higher under the controlled breathing protocol. Alternatively, our standard procedure for the controlled breathing protocol includes an initial period of unconstrained breathing to allow the individual's EELV to stabilise before commencing the washout. This is in line with current ATS/ERS recommendations [3] and may have ensured FRC differences were kept minimal between protocols.
Discrepancy in ventilation heterogeneity indices
While LCI – an index of global heterogeneity that is independent of SIII – was comparable between protocols (with a small bias), we found that SIII-dependent ventilation heterogeneity partitioned into proximal/conducting (Scond) and distal/acinar (Sacin) zones was not. Scond was poorly correlated between protocols, perhaps driven by a large between-session variability. Sacin was significantly different between protocols; moreover, Sacin obtained by free breathing was higher than that obtained by controlled breathing in individuals with relatively small VT or high RR. Between-protocol differences in Sacin appear to be independent of age, height or lung size, i.e. a change in breathing pattern was the dominant driver of the discrepancy.
There are two possible explanations for this. First, the effect could be purely methodological: when breaths are too “shallow”, estimation of SIII is rendered invalid as the N2 expirogram either never reaches a plateau or attains a plateau that is too short for reliable curve fitting, which would subsequently affect the derivation of Sacin and Scond. Anecdotal evidence suggests this occurs frequently. Methods to more reliably estimate SIII exist [21] but are still ultimately dependent on the presence of distinct phase II and III regions in the expirogram and an adequate portion of phase III. More useful may be methods of quantifying ventilation heterogeneity that do not rely on SIII estimation [22, 23] – these may help not just in free breathing but also in pathologies where SIII is often difficult to define. The feasibility and clinical relevance of these methods should be a direction for future investigation.
Alternatively, the effect may be physiological: when patients breathe shallower and faster, the fast-emptying lung compartments will increase their N2 washout while the slow-emptying lung compartments will be unable to empty completely, thus increasing apparent heterogeneity estimated by Sacin and Scond. Indeed, the foundational studies on ventilation distribution clearly demonstrated an effect of breathing pattern (including VT) on MBNW indices [4]. Recently, Ratjen and colleagues [24] investigated the effects of altered VT on MBNW indices and found that shallow breathing (compared to “unrestricted” breathing) significantly increased LCI. The authors speculated that this was due to the effects of increased dead space to VT ratio on the determinants of LCI (both CEV and FRC). Notably, these investigators did not examine the impact on Scond or Sacin.
We found that, on average, VT was shallower, RR was faster and CEV higher in the free breathing protocol in our study participants. The change in CEV would be expected, since smaller breaths may reduce alveolar gas mixing efficiency and therefore increase the volume (and time, though this was not significant) required to reach end-of-test criteria. This likely drove the changes in LCI we observed – however, differences in LCI were unrelated to differences in CEV (supplementary figure S5). It should be noted that the differences seen in LCI were very small relative to its magnitude and unlikely to be clinically significant. Furthermore, we found a weak relationship between differences in Scond and differences in CEV; this was not seen with Sacin (supplementary figure S6). However, when plotted against differences in CEV/FRC (i.e. LCI, which corrects CEV for lung volume), the relationship with Sacin became significant. This supports the interpretation that the differences seen in Sacin were driven by changes in ventilation heterogeneity, rather than alterations to lung volume.
Variability of MBNW indices
We also report between-session variability over 2–10 weeks for both protocols, using the same commercially available device. This allowed us to compare between-protocol differences in MBNW indices against their short-term variability. Within-session CoV of FRC and LCI was similar to published values for the free breathing protocol [25], using the same device. However the between-session coefficient of repeatability (equal to 1.96×sd of the mean differences) of LCI, Scond and Sacin was greater in our study, which may reflect the different time between sessions (weeks/months versus days). There is a paucity of repeatability data in health, and previous reports may not be generalisable due to persistent between-device differences [17, 18, 26]. Nevertheless, the excellent between-session repeatability for FRC and LCI [25, 27–29] but poor repeatability in Sacin and Scond [25, 28] has been noted with the free breathing protocol by other investigators, regardless of device, tracer gas or disease. The higher between-session repeatability values seen in Scond and Sacin with the controlled breathing protocol are consistent with other studies [30, 31].
It is also worth noting that FRC and LCI are derived from successive cumulation of volume at each breath, i.e. they are integrated measures, which are more robust to noise. On the other hand, Scond and Sacin are estimated from slopes, i.e. differentiated measures, which are inherently susceptible to noise. Furthermore, unlike FRC and LCI, the values of Scond and Sacin are very close to zero, which may explain the very low between-session ICC values observed. These may be fundamental reasons for the greater variability seen in Scond and Sacin but not FRC and LCI, and this variability becomes even more pronounced when variability in tidal breathing is introduced. This higher variability may also drive the differences seen between the two protocols. We also note the potential for these differences to be further exaggerated in disease – this and the high between-session variability observed may limit the clinical utility of SIII indices derived from free breathing. These speculations need to be confirmed in further investigations.
Significance
Both MBNW breathing protocols are designed to measure ventilation heterogeneity during resting tidal breathing. The controlled breathing protocol was originally devised based on modelling studies in which convection-dependent and diffusion-convection interaction-dependent mechanisms of airflow in the lung (the basis of Scond and Sacin, respectively) were first described [32]. The controlled breathing protocol standardises the VT at which ventilation distribution is assessed, and also ensures that expiration occurs well into the alveolar plateau so that a reliable SIII estimation can be made. This standardisation potentially minimises variability between tests. However, this “artificial” pattern of breathing may distribute the ventilation differently in lungs of different sizes, i.e. in shorter versus taller people. These differences may then affect the measurements of MBNW indices such that they no longer reflect true resting or “natural” ventilation distribution. In contrast, under a free breathing protocol, breathing occurs at (or closer to) natural resting breathing pattern; thus, the MBNW indices obtained arguably better reflect the individual's native ventilation distribution during normal tidal breathing.
Our results lend weight to the idea that controlled breathing imposes a condition that is different to the individual's natural resting breathing pattern, and that MBNW indices derived from this protocol may not necessarily reflect their “native” ventilation heterogeneity. Specifically, breathing at 1 L tended to improve heterogeneity measured by Sacin in individuals whose VT during free breathing was <1 L, or whose RR during free breathing was faster than the prescribed 8–12 breaths per minute. So, which is the “better” test? We suggest that it is whichever provides the highest sensitivity in detecting disease or assessing treatment. The “better” test may even be disease-specific and warrants further investigation.
Since these differences are apparent despite the use of VT-corrected versions of Scond and Sacin, our study suggests that the linear VT correction currently employed for the free breathing protocol may not be adequate. Again, methods which assess ventilation heterogeneity without the need for estimating SIII, ideally in a manner that is independent of VT [22], may be useful here.
Limitations
Our study has several limitations to be taken into consideration. First, a number of the volunteers could be considered “trained” in lung function testing. However, it is unlikely that prior experience would influence the largely effort-independent MBNW test. Second, 4 out of 27 participants required coaching before or after free breathing testing due to insufficient expired volume for SIII estimation. Thus, there is a chance our results may actually underestimate the true discrepancy between the two protocols, particularly in shallow breathers [24] or in lung disease where increased dead space further complicates SIII estimation [33]. Third, our results only apply to a healthy population tested on a single MBNW device. There are likely to be differences between equipment [17, 18] and between populations (e.g. children, older adults, lung disease) that limit the generalisability of our results.
Conclusion
We have demonstrated that while controlled and free breathing protocols are equivalent in terms of FRC and LCI, phase III-derived MBNW indices of ventilation heterogeneity are lower (i.e. better) when measured by the controlled breathing method. This effect seems to be driven largely by deviations in breathing pattern between the two protocols, particularly in patients who either breath faster or shallower than the criteria imposed by the controlled breathing protocol. Future work needs to be carried out in diseased patients to see if these observations hold true. Our study sheds light on potential physiological mechanisms behind these differences, and the overall interpretation of ventilation heterogeneity measured by MBNW. In better characterising the difference between the two protocols, these findings also help facilitate the ongoing efforts to standardise MBNW as an emerging clinical test.
Supplementary material
Supplementary Material
Please note: supplementary material is not edited by the Editorial Office, and is uploaded as it has been supplied by the author.
Supplementary material 00435-2020.SUPPLEMENT
Acknowledgements
We wish to acknowledge the help of the study participants, who kindly volunteered their valuable time for the study.
Footnotes
This article has been revised according to the correction published in ERJ Open Res 2021; 7: 50435-2020 [https://10.1183/23120541.50435-2020]
This article has supplementary material available from openres.ersjournals.com.
Support statement: There were no direct financial sponsors for the submitted work. S. Milne is supported by the MITACS Accelerate programme.
Conflict of interest: B.M. Handley has nothing to disclose.
Conflict of interest: E. Jeagal has nothing to disclose.
Conflict of interest: R.E. Schoeffel has nothing to disclose.
Conflict of interest: T. Badal has nothing to disclose.
Conflict of interest: D.G. Chapman has nothing to disclose.
Conflict of interest: C.E. Farrow has nothing to disclose.
Conflict of interest: G.G. King reports fees for consultancy services (which include lectures and advisory board services), conference attendance support and unrestricted research grants from AstraZeneca, Boehringer Ingelheim, CycloPharm, GlaxoSmithKline, Novartis, Menarini and MundiPharma, and research grants and fellowships from the National Health & Medical Research Council and the Asthma Foundation, outside the submitted work.
Conflict of interest: P.D. Robinson has nothing to disclose.
Conflict of interest: S. Milne reports personal fees from Novartis and Boehringer Ingelheim outside the submitted work, and is supported by the MITACS Accelerate programme.
Conflict of interest: C. Thamrin has nothing to disclose.
- Received June 25, 2020.
- Accepted October 16, 2020.
- Copyright ©ERS 2021
This article is open access and distributed under the terms of the Creative Commons Attribution Non-Commercial Licence 4.0.