TABLE 1

Reponses of automatic continuous positive airway pressure (APAP) devices to obstructive sleep apnoea simulated by the bench test

APAP devicePmax cmH2Otmax minPmean cmH2OResidual AHI events per hBreathing normalisation
A117.9, 17.822.4, 19.014.6, 14.62.0, 2.0Yes , yes
A215.4, 15.640.3, 57.713.4, 13.52.0, 1.5Yes, yes
B10.5, 10.5130.6, 121.36.7, 7.974.5, 71.0No, no
C13.4, 13.928.0, 44.112.4, 12.23.0, 3.0Yes, yes
D10.7, 10.775.1, 103.79.7, 9.726.5, 32.5No, no
E10.4, 10.420.7, 18.910.2, 10.23.5, 2.5No, no
F11.9, 12.134.4, 36.010.1, 10.111.5, 13No, no
G10.5, 11.032.2, 83.29.9, 9.933, 26.5No, no
  • The two values for each variable correspond to the results obtained in the two test repetitions in each device. Pmax: maximum positive airway pressure applied; tmax: time to reach Pmax±0.3 cmH2O; Pmean: mean positive airway pressure; AHI: apnoea–hypopnoea index; A1: AirSense 10, standard setting; A2: AirSense 10, response setting; B: Dreamstar; C: Icon; D: Resmart; E: Somnobalance; F: System One; G: XT-Auto.