Differences in in vitro functional residual capacity (FRC) measurements from the lung model functional residual capacity for four different lung volumes
Lung model FRC volume | Subjects | WIMR | EM | ndd v.2.00 | ndd v.2.01 | ||||
Absolute | Relative | Absolute | Relative | Absolute | Relative | Absolute | Relative | ||
2.31 | 9 | −0.01±0.02 | −0.2±0.8% | 0.07±0.02 | 3.3±1.0% | −0.54±0.14 | −23.4±6.1% | −0.15±0.04 | −6.3±1.7% |
2.11 | 9 | −0.01±0.02 | −0.43±0.9% | 0.06±0.02 | 2.9±0.9% | −0.39±0.12 | −18.5±5.7% | −0.11±0.03 | −5.2±1.5% |
1.81 | 9 | 0.03±0.02 | 1.6±1.4% | 0.07±0.02 | 3.7±1.0% | −0.44±0.11 | −24.1±6.4% | −0.09±0.03 | −4.7±1.6% |
1.51 | 9 | 0.02±0.03 | 1.7±13.1% | 0.02±0.02 | 1.1±1.3% | −0.26±0.05 | −17.1±3.0% | −0.08±0.02 | −5.0±1.2% |
Data presented as mean±sd, unless otherwise stated. Absolute measurements are in litres. WIMR: Woolcock in-house device; EM: Exhalyzer D device; ndd v.2.00: EasyOne Pro LAB device Version 2.00.01.05; ndd v.2.01: EasyOne Pro LAB device analysed in Version 2.01.00.09.