In vitro functional residual capacity (FRC) measurements on each device
Lung model FRC volume | Trials | WIMR | EM | ndd v.2.00 | ndd v.2.01 |
2.31 | 9 | 2.30±0.02 | 2.39±0.02 | 1.78±0.14 | 2.16±0.04 |
2.11 | 9 | 2.10±0.02 | 2.17±0.02 | 1.72±0.12 | 2.00±0.03 |
1.81 | 9 | 1.84±0.02 | 1.88±0.02 | 1.37±0.11 | 1.72±0.03 |
1.51 | 9 | 1.53±0.03 | 1.53±0.02 | 1.25±0.05 | 1.43±0.02 |
Data presented as mean±sd, unless otherwise stated. Measurements are in litres. WIMR: Woolcock in-house device; EM: Exhalyzer D device; ndd v.2.00: EasyOne Pro LAB device Version 2.00.01.05; ndd v.2.01: EasyOne Pro LAB device analysed in Version 2.01.00.09.