TABLE 1

Exhaled carbon dioxide detected with four machines and three circuit configurations

CircuitBreas Vivo 50Breas NIPPY 3+Philips Respironics A40ResMed AirSense CPAP
Exhalation port2.93.13.23.7
Active valve
 0 cmH2O PEEP2.11.6
 4 cmH2O PEEP1.5
Vented mask2.72.82.73.6

Data are given in kPa. CPAP: continuous positive airway pressure; PEEP: positive end-expiratory pressure.