Abstract

Background: Electronic monitoring (EM) is used increasingly to measure medication non-adherence. Unbiased EM assessment requires fulfillment of assumptions. The purpose of this study was to determine assumptions needed for internal and external validity of EM measurement. To test internal validity, we examined if (1) EM equipment functioned correctly, (2) if all EM bottle openings corresponded to actual drug intake, and (3) if EM did not influence a patient's normal adherence behavior. To assess external validity, we examined if there were indications that using EM affected the sample representativeness.

Methods: We used data from the Supporting Medication Adherence in Renal Transplantation (SMART) study, which included 250 adult renal transplant patients whose adherence to immunosuppressive drugs was measured during 3 months with the Medication Event Monitoring System (MEMS). Internal validity was determined by assessing the prevalence of nonfunctioning EM systems, the prevalence of patient-reported discrepancies between cap openings and actual intakes (using contemporaneous notes and interview at the end of the study), and by exploring whether adherence was initially uncharacteristically high and decreased over time (an indication of a possible EM intervention effect). Sample representativeness was examined by screening for differences between participants and non-participants or drop outs on non-adherence.

Results: Our analysis revealed that some assumptions were not fulfilled: 1) one cap malfunctioned (0.4%), 2) self-reported mismatches between bottle openings and actual drug intake occurred in 62% of the patients (n = 155), and 3) adherence decreased over the first 5 weeks of the monitoring, indicating that EM had a waning intervention effect.

Conclusion: The validity assumptions presented in this article should be checked in future studies using EM as a measure of medication non-adherence.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Adult
  • Algorithms
  • Drug Monitoring / instrumentation
  • Drug Monitoring / standards*
  • Drug Packaging / instrumentation
  • Female
  • Humans
  • Immunosuppressive Agents / therapeutic use*
  • Interviews as Topic
  • Kidney Transplantation*
  • Male
  • Middle Aged
  • Patient Compliance / statistics & numerical data*
  • Reproducibility of Results
  • Surveys and Questionnaires
  • Switzerland

Substances

  • Immunosuppressive Agents